Fixture Calibration
Step 1: Measure connector delay
- Solder disk to end of connector
- Load Female Arbitrary Calibration Settings
SDR-Kit BNC Calibration Kit
- S11 OSL calibration using Female Calibration Devices
- Measure Connector
- Adjust Port Extension for flat S11 ImagZ and S11 Phase
- Multiply result by two, 41ps * 2 = 82ps.
|
Shorted Connector and Calibration Devices
VNWA results with Port Extension
Step 2: Load and modify Male Calibration Settings
- Load Male Arbitrary Calibration Settings
SDR-Kit BNC Calibration Kit
- Add -82ps to OPEN SHORT LOAD Delay values
- Set THRU Transmission Delay value to zero
|
Initial OPEN Calibration Settings
Modified OPEN Calibration Settings
Step 3: Fixture S11 OSL Calibration
- Connect Fixture to Port 1
- OSL Calibration with Cal devices on Port 2 side of fixture
|
S11 Calibration
Step 4: Fixture Thru and Thru Match Calibration
- Connect Port 2 to Fixture
- Thru and Thru Match calibration
|
Thru and Thru Match Calibration
Step 5: Check Calibration
- 12-term correction measurement using F2 Key
- Evaluate measurements
- t2s(S21) RealZ slightly off
|
VNWA Results
Step 6: Refine Calibration
VNWA Results
Measure 3.3kΩ 1/4W Resistor
VNWA Files: ResistorMeas.zip
DC Measured = 3.237kΩ
#22 Calibration Wire
#24 Calibration Wire
SimSmith Model
SS-3k3.ssx
Measure 1.5kΩ 1% Metal Film 1/4W Resistor
Resistor Calibration Wire
Bridge wire parallel to resistor bottom horizontal plane
VNWA Resistor Calibration Wire
SimSmith Model
SS-1k5.ssx
Measure 1.5kΩ Carbon Film 1/4W Resistor
Resistor DUT
VNWA Resistor Calibration Wire
SimSmith Model
SS-1k5-CF.ssx